GDS OTV Insert
SIAL/73021BC - BioControl, For use with GDS, Performance verification of the GDS Rotor-Gene
Product Type: Equipment/Book
application(s) | food and beverages pathogen testing |
compatibility | for use with GDS Rotor-Gene |
manufacturer/tradename | BioControl Systems 73021 |
shelf life | 3‑5 mo. |
storage condition | protect from light at room temperature |
storage temp. | room temp |
technique(s) | PCR: suitable |
Compatibility: | BioControl GDS |
Components: | • OTV Insert – good for 30 verification runs • Optical Filters for Rotor-Gene 3000 and Rotor-Gene Q, one each |
General description: | The GDS OTV (Optical Temperature Verification) Insert provides automated performance verification and dynamic calibration of the GDS Rotor-Gene. The OTV system uses the chemical properties of thermochromic liquid crystals (TLC) as an absolute temperature reference. When heated, a TLC changes from opaque to transparent at a very precise temperature. The Rotor-Gene can measure this precise temperature transition point and, by comparing the measured in-tube thermal transition point to the known calibrated value, the OTV software verifies that the Rotor-Gene is within specification. If necessary, the OTV software allows the user to automatically recalibrate the instrument with a simple push of a button. A printable post-run report documents each test and any necessary recalibration. |
Testing Method: | Optical Temperature Verification |
Storage Temp. | room temp |
UNSPSC | 41116135 |