Reference Filter Set
SIGMA/EP6135928001 - for Eppendorf BioSpectrometer®
Product Type: Equipment/Book
| feature | for Eppendorf BioSpectrometer® |
| manufacturer/tradename | Eppendorf® 6135928001 (Global) |
| Features and Benefits: | • Filter traceable to NIST • Filter recertification via Eppendorf (contact: Eppendorf Service) • Inspection protocol preprogrammed in device • Verification of wavelength accuracy and precision at 260 nm, 280 nm and 800 nm • Verification of the photometric accuracy and precision in the 260 nm to 800 nm range • The variation coefficient is listed along with the average value of 15 measurements. • Contains 7 filters (1 blank filter, 3 filters for checking UV/Vis wavelenght systematic error and precision, 3 filters for photometric verification) |
| General description: | Contains 7 filters (1 blank filter, 3 filters for checking UV/Vis wavelenght systematic error and precision, 3 filters for photometric verification) |
| General description: | The secondary UV/Vis Eppendorf BioSpectrometer Reference Filter Set is used to verify photometric accuracy and wavelength systematic error. |
| Legal Information: | Eppendorf BioSpectrometer is a registered trademark of Eppendorf AG |
| Legal Information: | Eppendorf is a registered trademark of Eppendorf AG |
| UNSPSC | 41104900 |