Merlin Microseal™ System Replacement Septum
SUPELCO/24818-U - 1 SPME septum
Synonym: GC accessory, SPME septum, Merlin Microseal
Product Type: Equipment/Book
Application: | Merlin Microseal™ System may be used to replace traditional septum to avoid leaking and bleeding caused during operating GC/MS system. It may also be equipped to the injection port during GC analysis for determination of plasticizer N-butylbenzenesulfonamide and the pharmaceutical Ibuprofen in wastewater. |
General description: | The Merlin Microseal System is a septum-less system that provides very long life. Because there is not a septum to pierce, septum fragments will not be formed and deposited in the inlet liner. The septum contains a series of seals (wiper rib, o-rings, and a duckbill valve) that allow a needle to enter while maintaining a leak-free seal. The septum can be used only with a syringe that has a 23 gauge "blunt tipped" needle, or with an SPME fiber assembly with a 23 gauge needle. The thicker shaft is required to make the necessary contact with the septum seals. |
Legal Information: | MicroSeal is a trademark of Merlin Instrument Co. |
Other Notes: | Three septa versions are available: • Low Pressure for use with 23 gauge syringe needles, and injection port pressures between 1 and 45 psi. Do not use with syringe needles that have beveled tips. • General Purpose for use with 23 gauge syringe needles, and injection port pressures between 3 and 100 psi. Do not use with syringe needles that have beveled tips. • SPME for use with 23 gauge SPME fiber assemblies. Do not use with SPME fiber assemblies that have beveled tips. Note: Do not use with beveled tips. |
RIDADR | NONH for all modes of transport |
WGK Germany | WGK 3 |
Flash Point(F) | Not applicable |
Flash Point(C) | Not applicable |
UNSPSC | 12000000 |